Cytiva (Akta Pure 25 L1)
  • System pump
    • Pump type: piston pump
    • Flow rate: 0.0010 25 mL/min (up to 50 mL/min during column packing)
    • Pressure range: 0 – 20 MPa
  • Sample pump
    • Pump type: piston pump
    • Flow rate: 0.001 – 50 mL/min
    • Pressure range: 0 – 10 MPa
  • Mixer
    • Mixing principle: chamber with a magnetic stirrer
    • Mixer volume: 5 mL
    • Gradient flow rate range: 0.1 – 25 mL/min
  • UV monitor
    • Wavelength range: 280 nm
    • Resolution: 0.001 mAU
    • Linearity:  ± 5% within 0–2 AU
    • Drift:  ≤ 0.2 mAU; AU/h, 2 mm cell ≤ 0.2 mAU; AU/h at 280 nm, 2 mm cell
  • Conductivity monitor
    • Conductivity reading range: 0.01 mS/cm – 999.99 mS/cm
    • Accuracy:  ± 0.01 mS/cm or ± 2%, whichever is greater (within 0.3 to 300 mS/cm)
    • Operating pressure: 0 – 5 MPa
  • Fraction collector, F9-R
    • Number of fractions: 175 fractions
    • Vessels: 175 x 3 mL tubes, 85 x 8/15 mL tubes
    • Fraction volumes: 0.1 – 50 mL
  • Sample Condition: Wet sample in buffer solution
  • pH: known compatibility with planned column to be used
  • Solvent/Buffers known compatibility with planned column to be used, must be filtered and degassed (sonicated)